OverviewSpecificationsResources

Dynamic Reducibility DR 5000, DR 2000

DR 5000 system is designed to operate dynamic test according to ISO 13930 (Dynamic test for low-temperature reduction-disintegration – LTD)

DR 2000 system is designed to perform both Determination of the low-temperature reduction (according to ISO 11257) and Reduction Degradation (according to IS 10823 )

The equipment is based on an electric horizontal oven including two half cylindrical heating elements where the rotating reaction tube is placed.

The Control Cabinet provides for temperature, gases flow rates and tube rotation motor control.

From experiences of several laboratories involved in these tests, it has been pointed out that results reliability greatly depends on temperature and reaction gas flow rate accuracy, therefore electronic mass flow meter and high quality temperature controller are supplied.

DR 5000DR 2000
ISO 13930 ISO 11257 IS 10823 Sec 2
LTD Reduction Degradation
OVEN and CONTROL UNIT
Max Operation power 7 kW 4 kW
Flow rate measuring device Digital Mass Flow Controller
ROTATING RETORT
Vessel Length 540 mm 200 mm
Vessel Diameter150 mm 130 ± 1 mm 130 mm
Resolution 4 ———– 4
TEST DATA
Sample Mass 500 g ± the mass of 1 particle 500 ± 1 g
Test Temp500 ± 5 °C 760 ± 5 °C 550 ± 10 °C
Reduction Time 60 min 300 min 60 min
Reaction Gas Mix Mix A Mix E Mix B
Reaction Gas Flow rate 20 ± 1 nL/min13 ± 0.5 nL/min 15 nL/min
Rotation speed10 .± 0.2 RPM10 ± 1.0 RPM10 ± 0.5 RPM
MAIN POWER: 220 V + Neutral (7 kW max)220 V + Neutral (4 kW max)