A Product by Swan Analytical Instruments
OverviewConductivity MeasurementSpecificationsResources
Online Conductivity / Resistivity Anlayzer having monitor system for the automatic, continuous measurement of the specific (total) conductivity in feedwater, steam and condensate.
  • Transmitter AMI Powercon in a rugged aluminum enclosure (IP 66).
  • Swansensor UP-Con1000-SL two-electrode conductivity sensor with slotlock design andintegrated Pt1000 temperature probe.
  • Flow cell QV-Flow UP-CON-SL made of stainless steel with flow adjustment valve and digital sample flow meter. Quick sensor release with patented slotlockdesign.
  • Conductivity measurement range: 0.055 to 1000 μS/cm
  • Big backlit LC display for the reading of measuring value, sample temperature, sample flow, temperature compensation type and operating status.
  • Data logger for 1’500 data records stored at a selectable interval. (Data download to PC
Swansensor UP-Con1000-SL with integrated Pt1000 temperature probe (k = 0.0415 cm-1).
Measuring rangeResolution
0.055 to 0.999 μS/cm0.001 μS/cm
1.00 to 9.99 μS/cm0.01 μS/cm
10.0 to 99.9 μS/cm0.1 μS/cmH
100 to 1000 μS/cm1 μS/cm
Automatic range switching.
Accuracy:± 1 % of measured value
Temperature compensations
automatic, according to:
- Non linear function (NLF) for high purity water
- Neutral salts
- Strong acids
- Strong bases
- Ammonia, Ethanolamine
- Morpholine
- Linear coefficient in %/°C
- Absolute (none)
Temperature measurement with Pt1000 type sensor
Measuring range:-30 to +130 °C
Resolution:0.1 °C
Sample flow measurement
with digital SWAN sample flow sensor.
Transmitter Specifications and Functionality
Electronics case:Cast aluminum
Protection degree:IP 66 / NEMA 4X
Display:backlit LCD, 75 x 45 mm
Electrical connectors:screw clamps
Voltage:100 - 240 VAC (± 10 %), 50/60 Hz (± 5 %)
or 24 VDC, (± 15 %)
Power consumption:max. 20 VA

Monitor Data


Sample conditions
Flow rate:5 to 20 L/h
Temperature:up to 50 °C
Inlet pressure:up to 2 bar
Outlet pressure:pressure free
Flow cell and connections
Flow cell made of stainless steel with built-in flow adjustment valve and digital sample flow meter. Quick sensor release with patented slot-lock design.
Sample inlet:Swagelok ¼” tube adapter
Sample outlet:G ½” adapter for flexible tube Ø 20 x 15 mm
Panel
Dimensions:850 x 280 x 200 mm
Material:stainless steel
Total weight:12.0 kg
Resources